The use of gel imaging analysis system and points for attention
Gel imaging analysis system: it can be applied to protein electrophoresis gel, DNA gel, sample collection and qualitative and quantitative analysis. Samples include: EB, SYBR Green, SYBR Gold, Texas Red, GelStar, Fluoroscecin, Radiant and so on to monitor nucleic acid monitoring. And Coomassie Blue, SYPRO Orange, various stained protein gels, such as Kao dye. Or UV, EB and color and visible sample imaging)
Precautions for use:
Pay attention to the boot sequence, first open the gel imaging system, then turn on the computer to enter the software.
When UV Gel is taken, the EB contamination instrument should be prevented. The door of the gel imaging system can not be exposed to contaminated gloves, so that gloves can not be contaminated when software is operated.
• the front panel of the gel imaging system can not be opened in the process of using UV source photography.
After taking photos, take out the waste glue and wipe it with soft paper.
Focus gently and do not move violently.
When the ambient voltage is unstable, please use a regulated power supply. In case of power failure during use, please turn off the power supply of the instrument in time until the power comes back.
When using, please turn on the power switch of the instrument first, then turn on the computer switch and turn on the software (quantity one).
Keep the environment in the observation room dry and wipe the water or other liquid left on the observation board in time (soft paper can be used, generally roll paper).
When observing the gel used in EB dyeing, be careful not to pollute the surface of the instrument. Never contact your gel directly with your hands or contact gloves with contacts to touch the door of the instrument and the handle of the observation table.
When using the instrument, the door and the observation platform should be closed tightly, otherwise the ultraviolet lamp will not be used normally.
As far as possible, do not connect the computer to the Internet or local area network, and install anti-virus software on the computer (genuine software is recommended), so as to make the computer special.
When the instrument is not used for a long time, please cover it with dust cover.
In order to prolong the service life of the tube, please observe the gel well and turn off the light source in time. (the instrument itself has a 10 minute automatic protection program).